How to Prepared and Standardize of 0.1 N Ammonium Thiocynate Volumetric Solution
Preparation of 0.1 Ammonium Thiocynate Volumetric solution
➤ Glass ware and instruments required.
Beaker, Glass stirrer, Graduated Pipettes, Graduated cylinder, Analytical balance, Sucker, Volumetric flask, Burette, Conical Flask, Fume hood.
Beaker, Glass stirrer, Graduated Pipettes, Graduated cylinder, Analytical balance, Sucker, Volumetric flask, Burette, Conical Flask, Fume hood.
➤ Reagents required.
Ammonium Thiocynate Reagent Grade
Molecular Wt. :- NH4SCN,76.12
Silver nitrate Volumetric Solution (0.1 N),
Nitric Acid,
Ferric Ammonium Sulfate Solution.
Ferric Ammonium Sulfate Solution.
➤ Preparation.
Weigh about 8 g of ammonium thiocynate reagent grade in 1000 ml volumetric flask. Dissolve and dilute with water. Filter the
solution, if it is not clear.
➤ Standardization.
Pipette out 25 ml of standard 0.1
N silver nitrate VS into glass stoppered flask. Dilute with 50 ml of water,
then add 2 ml of nitric acid and 2 ml of ferric ammonium sulphate solution, and
titrate with the ammonium thiocynate solution to the first appearance of red
brown colour. Note this final end point as burette reading in ml.
➤ Calculations.
ml AgNO3 x N /
M AgNO3
N / M =
----------------------------------------------------------------------
ml NH4SCN solution [ Burette reading in ml ]